32 of 82
3) Press the Return icon to save changes and return to the main screen.
Changing the Needles
3.9
The 9705 comes with two vent needles to allow sampling from a variety of bottle sizes.
• The shorter vent needle is used to sample 5-30 mL bottles.
• The longer vent needle is used to sample 50-125 mL bottles.
The vent needle must be removed to use the Large Volume Adapter (LVA); this process is
described in
Note:
The short sampling needle does not need to be removed at any time. Do not attempt to
remove the shortest needle.
To change needles:
1) Tilt the bottle dock forward.
2) Remove the bottle guide.
Operator Manual
"Connecting the Large Volume Adapter" on page
Fig 3-10 : Tilt Bottle Dock
System Setup - 9705
35.
HIAC