Section 8 Troubleshooting - Hach LANGE LICO 500 User Manual

Table of Contents

Advertisement

Section 8
Problem/Display screen
Please insert adapter Z.
Color = ***
Absorbance > 3.5!
Concentration too high!
Error
Selfcheck stopped.
Please check the lamp.
Please close the lid.
Error [xx]
Error
Selfcheck stopped. Please remove
the cuvette
Please close the lid.
Error
Selfcheck stopped.
Hardware error.
Error [x]
Error
Too much ambient light!
Move device into shade
or close the lid!
Negative result!
No evaluation!
Over measuring range
Under measuring range
Troubleshooting
Likely Cause
For measurements with 11mm round
cells the adapter Z is required.
Spectrum locus outside the measuring
range.
The measured absorbance exceeds 3.5
Calculated concentration is higher than
999999
Self-Check Test stops while starting the
instrument
Self-Check Test stops while starting the
instrument
Electronic defect
The instrument sensors detects too
much ambient light.
The calculated result is negative
Error in the test database / user
database
The measured absorbance is above the
calibration range of the test
The measured absorbance is below the
calibration range of the test
Action
Insert the adapter Z into cell
compartment #2.
Press OK.
Dilute the sample or select appropriate
color scale.
Dilute the sample and repeat the
measurement
Dilute the sample and repeat the
measurement
Check the lamp and replace, if
necessary.
Close the lid.
Press Start Again.
Remove the cuvette/sample cell from
the cell compartment.
Press OK.
Contact the manufacturer or a sales
representative and indicate the error
number
Reduce ambient light. (Avoid direct sun
light.)
Close the lid.
Check the concentration of the sample
Check the programming
Contact the manufacturer or a sales
representative
Dilute the sample and repeat the
measurement
If possible, select a test with a lower
measurement range or use a
cuvette/sample cell with a longer path
length
91