Setting the storage
interval
4.4.2 Switching on AutoStore (Int 1)
The storage interval (Int 1) determines the time interval
between automatic storage processes.
After the time interval expires, the current data record is
transmitted to the data storage and to the printer/interface.
The storage interval (Int 1) is set to OFF in the factory.
Thus, the AutoStore function is switched off.
To switch the function on, set up a time interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min).
Note
When the AutoStore function is is switched on, the setting of
the data transmission interval (Int 2) is ineffective (see page
110).
1 Press and hold down the
s
2 Press the
u d
3 Press the
interval between storage processes.
4 Confirm with
The number of free memory locations appears on the
display.
g
key.
key. Int 1 appears on the display.
keys to set up the required time
g
.