3. Installation and start-up

3.1 Installation
When unpacking, check that no parts are damaged. If any damage is found, contact Care of
Sweden before using the product.
1. Place the mattress on top of an existing mattress. If possible, secure it by pulling the elastic
straps on the bottom over the corners of the underlying mattress.
Note: If the mattress has an integrated foam core
mattress directly on the bed base. Secure the mattress to the moving parts of the bed base
using the straps on the cover.
2. Hang the pump by the foot end of the bed or place it on a flat, sturdy surface. Place the
power cord so there is no risk of stumbling over it, run it over with the bed wheels or getting it
jammed when raising or lowering the bed.
3. Connect the air tubes coming from the mattress to the side of the pump. The pump and
hoses are equipped with quick connectors; an audible "click" will be heard when these have
been connected correctly.
4. Check to make sure that the power is set to "0" (off). Plug the power cord into an
approved outlet (110 or 220 V depending on the country and the voltage level).
NOTE:
• Make up the bed with sheets to increase user comfort.
• Check that the CPR valve is closed.
• Make sure that the mattress is the correct size for the bed.
3.2 Start-up and settings
PLEASE NOTE:
1. Start the pump by pressing "1" (on), on the side of the pump.
2. The pump starts filling the mattress with air. The green light with the
during the inflation phase; it takes about 15 minutes for the mattress to fill with air.
3. When the
symbol shows a steady green light, the mattress is ready to be used.
3.3 Comfort settings
The air pressure in the mattress can be adjusted according to the user's comfort preference.
Always consult the prescriber or other authorised personnel for the appropriate setting.
Appropriate air pressure is selected using the knob (0–250 kg). The numbered scale guides
your selection. For sitting positions, we recommend that the pressure be raised by +25
Check using the hand check (see section 3.4).
Always read the user manual prior to use.
(CuroCell