Care of Sweden CuroCell 4 User Manual

Care of Sweden CuroCell 4 User Manual

Curocell 4 alternating pressure system

Advertisement

User manual
2019-04-26
CuroCell
4
®
Alternating pressure system
Item number user manual: 95-001044

Advertisement

Table of Contents
loading

Summary of Contents for Care of Sweden CuroCell 4

  • Page 1 User manual 2019-04-26 CuroCell ® Alternating pressure system Item number user manual: 95-001044...
  • Page 2: Table Of Contents

    Contents Important safety precautions Introduction Product description Assembly/Installation Operation Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information...
  • Page 3: Important Safety Precautions

    Care of Sweden. 2. Do not use the product in wet rooms or where there is a risk of the pump coming into contact with water or another liquid.
  • Page 4: Introduction

    Care of Sweden. The system has been tested and approved according to the following standards: IEC 600601-1, IEC 60601-1-2, SS-EN ISO 12182, SS-EN ISO 14971, SS-EN ISO 10993, EN 597-1, EN 597-2, EN 14126.
  • Page 5 NOTE! • For certain patients, such as amputees, are the recommended minimum weight not valid. These users may however have other requirements as the entire mattress is not being loaded. Please refer to sections 4.2 and 4.3 for operating control. •...
  • Page 6: Product Description

    2. Product description 2.1 Pump and mattress 1. CPR (rapid deflation) a) Head b) Foot 2. Pump 3. Mattress 4. Pump connector 1a) Tubing system with CPR in head 4 static head cells 9 ventilated cells (optional) Static safety cells (Cell-in-cell) 1b) Tubing system with CPR in foot 4 static head cells 9 ventilated cells (optional)
  • Page 7 2.3 Pump functions 1. Power on/off (1/0) ON/OFF Button to turn the power on or off. 2. Lamp for normal pressure NORMAL PRESSURE At start-up: Lights when the mattress has reached the set pressure and is ready for use. During operation: Steady green light while pressure is normal.
  • Page 8: Assembly/Installation

    3. Assembly/Installation When unpacking the mattress, check that none of the parts are damaged. In the event of damage, contact the manufacturer or your local dealer before using the product. 1. Place the mattress on the bed base; use the straps on cover to securely fasten the mattress to the bed base.
  • Page 9: Operation

    4. Operation Note! Always read the user manual prior to use. 4.1 Start-up 1. Start the pump by pressing 1 (On). 2. Air starts to inflate the mattress. Note! We recommend that the comfort dial be set to 8 (max) when inflating a mattress.
  • Page 10 4.4 Alarms When the pressure in the mattress drops to an abnormally low level, the yellow low pressure lamp lights and an acoustic alarm sounds. First, check that the CPR valve is closed and that all connections are intact and correctly installed in accordance with paragraph 3.
  • Page 11: Cleaning And Reconditioning

    4.7 Heel function (with check valves, option) This feature allows pressure relief of the heels. Disconnect the air cell which is directly under the heel. This air cell will then be emptied for air. Press the bottom (1) and disconnect the plug. Let the cell remain in its place in the mattress.
  • Page 12: Storage

    INNER COVER AND COMFORT FABRIC (accessories) Clean the affected area with a mild cleaning solution for this purpose such as: alcohol with or without tensides, Isopropanol 70 %, oxidating solutions or chlorine max 1 %. 5.2 Reconditioning PUMP See section 5.1 for instructions. MATTRESS 1.
  • Page 13: Maintenance

    7. Maintenance 7.1 General 1. Check the power cord and plug for damage and scratches. 2. Check that the mattress cover is intact and that the cover and air cells are correctly assembled. 3. Check that the air flow alternates between each outlet on the side of the pump. 4.
  • Page 14: Troubleshooting

    Can be resolved by putting the pump on a flat, steady surface or by placing a towel between the pump and bed. If your questions are not answered in the above information, please contact Care of Sweden or your local dealer.
  • Page 15: Technical Description

    80/85/90/105/120x200/210x21 cm Weight 10.5 kg (80x200 cm) NOTE: Care of Sweden retains the right to modify the product specification at any time. WARNING! Follow operating and maintenance instructions carefully to prevent injury, as well as damage to the product or other equipment.
  • Page 16 9.3 Symbol Definitions XXXX-XX-XX Article No Manufacturer Year-Month-Day Foot placement Read the user manual CE marked acc. to Group of function Recycling MDD 93/42/EEC Recommended user weight Not in domestic refuse, Patient information Directly on bed base follow recycling instructions –...
  • Page 17: Spare Parts

    90x21 cm CC-901333 Heel cell, PU 105x21 cm CC-901334 Heel cell, PU 120x21 cm CC-8217 Fuse (set of 2) CC-8201 Filter, washable If other spare parts are needed, please contact Care of Sweden or your local dealer for further assistance.
  • Page 18: Options And Accessories

    11.3 Other accessories To ensure the best level of comfort, we recommend use of accessories from Care of Sweden since our products are designed to work well together. Please contact Care of Sweden or your local dealer for accessories or more information on our products.
  • Page 19: Warranty

    Service and maintenance must always be performed by Care of Sweden or one of its authorised technicians. Only use spare parts that have been approved by Care of Sweden. Use of non-approved spare parts invalidates the warranty.
  • Page 20 SE-514 23 Tranemo SWEDEN Visit address: Fabriksgatan 5A SE-514 33 Tranemo SWEDEN Cargo address: Byns väg 4A SE-514 33 Tranemo SWEDEN Supplied by Care of Sweden and CuroCell are trademarks of Care of Sweden AB © Care of Sweden AB, 2019...