3. Assembly/Installation
When unpacking the mattress, check that none of the parts are damaged. In the event of
damage, contact the manufacturer or your local dealer before using the product.
= Mode 1 = Alternating mode, every other cell is inflated (not the three
static head cells). The pressure change between cells, a cycle is ten
minutes. We recommend that this mode is used by default.
= Mode 0 = Static mode, the same pressure in all cells. When the Static
button is pressed, the alternating process is terminated and the pressure
between the cells evens out. The function can be used in association
with patient care,
repositioning/transferring in and out of bed, or as a
treatment mode.
Power off/on (1/0).
Suitable air pressure is set with the comfort dial (1‐8). Guidelines based
on user weight is given in the table on the pump. NOTE! In all uses
where the head end of the bed is raised,
at least one (1) step.
Lights when the mattress is ready for use.
Alarms for low pressure. In order to solve it see chapter 8,
troubleshooting.
1. Place the mattress on the existing mattress and fasten it
with the elastic bands around the underlying mattress.
If the mattress has a built‐in foam pad (CuroCell® Nova
CX19), it can be placed directly on the bed frame. Secure
the mattress to the bed base with straps on the cover.
Note!
• Make the bed with sheets for added comfort.
• Check that the CPR valve is closed.
• The mattress must be attached to the bed's moving
parts, if it is height adjustable (CuroCell® Nova CX19).
• Check that all the snaps in each cell and cover are
fastened correctly.
the pressure must be increased
7