Care of Sweden CuroCell S.A.M. User Manual

Care of Sweden CuroCell S.A.M. User Manual

Static antidecubitus mattress
Hide thumbs Also See for CuroCell S.A.M.:

Advertisement

User manual
2018-12-21
CuroCell S.A.M.
®
Static Antidecubitus Mattress
User manual item number: 95-001298

Advertisement

Table of Contents
loading

Summary of Contents for Care of Sweden CuroCell S.A.M.

  • Page 1 User manual 2018-12-21 CuroCell S.A.M. ® Static Antidecubitus Mattress User manual item number: 95-001298...
  • Page 2: Table Of Contents

    Contents Important safety instructions Introduction Product description Installation and start-up Cleaning and reconditioning Storage Maintenance Troubleshooting Technical description Spare parts Options and accessories Warranty Other information Own notes...
  • Page 3: Important Safety Instructions

    Care of Sweden. Do not use the pump in wet rooms or where there is a risk of the pump coming into contact with water or another liquid.
  • Page 4: Introduction

    We would be very grateful for all information involving accidents or incidents relating to our products, by reporting them immediately to us, here at Care of Sweden. This system has been tested and...
  • Page 5: Product Description

    1.3 Other information Action plan Pressure ulcers To achieve the best possible result for this product it is recommended that a well-planned and well-structured working environment be used. In our guideline "Action plan" you will find examples which can be applied and put into practice when using our products. Mattress guide This practical guide gives you as a care giver, help in choosing the correct mattress from the Care of Swedens range.
  • Page 6: Installation And Start-Up

    When unpacking the mattress, check that none of the parts are damaged. In the event of damage, contact Care of Sweden or your local dealer before using the product. 1. Place the mattress on top of an existing mattress. If possible, secure it by pulling the elastic bands on the bottom over the corners of the underlying mattress.
  • Page 7 2. Hang the pump by the foot end of the bed, or place it on a flat sturdy surface. Place the power cord so there is no risk of stumbling over it, run it over with the bed wheels, or getting it jammed when raising or lowering the bed.
  • Page 8 3.4 Hand check Hand check is used to ensure that the mattress is properly inflated. This should be done regularly, once per shift, after changing the position in the bed or changing the comfort settings. Open the cover. Place a flat hand under the air cells directly underneath the user’s pelvic region and make sure that the user does not lay directly on the underlying mattress.
  • Page 9: Cleaning And Reconditioning

    4. Cleaning and reconditioning It is important to follow the instructions below before using the product for another user. Consult the hygiene responsible or the supplier for help and instructions when unsure. 4.1 Cleaning PUMP VARNING! Do not open the pump housing – risk of electric shock. Service and maintenance must be performed by authorised technicians.
  • Page 10: Storage

    4.2 Reconditioning PUMP See section 4.1 for instructions. MATTRESS 1. Disconnect the tubes and let the air out, also open the CPR valve. 2. Remove the cover. See instructions in chapter 4.1 for cleaning covers. If the product is equipped with a foam core this should be removed and cleaned separately. 3.
  • Page 11: Maintenance

    Pump bag (accessory) can also be used. If your questions are not answered by the above information, please contact Care of Sweden or your local dealer for further assistance.
  • Page 12: Technical Description

    8. Technical description 8.1 Technical specification PUMP SPECIFICATION Model CuroCell S.A.M.® Voltage 220 V-240 V, 50 Hz Current consumption 0.05 A Fuse type T1AL, 250 V Dimensions (LxWxH) 25 x 12.5 x 8.5 cm Weight 1.4 kg Surroundings Temperature Operations: 10°C 40°C ˜...
  • Page 13 8.3 Symbol Definitions XXXX-XX-XX Article No Manufacturer Year-Month-Day Foot placement Read the user manual Group of function CE marked acc. to Recycling MDD 93/42/EEC Recommended user weight Patient information Not in domestic refuse, Directly on bed base – category follow recycling instructions Place on top of Counteract shear force Do not wash...
  • Page 14: Spare Parts

    120x210x10 cm Art.No Description CC-250101-E3 CuroCell S.A.M.®, pump without quick connection CC-250102 CuroCell S.A.M.®, pump incl. quick connection CC-908032 Tubing set incl, CPR If other spare parts are needed, please contact Care of Sweden or your local dealer for further assistance.
  • Page 15: Options And Accessories

    Pump bag, black 10.3 Other accessories To ensure the best level of comfort, we recommend use of accessories from Care of Sweden since our products are designed to work well together. Please contact Care of Sweden or your local dealer for accessories or more information on our products.
  • Page 16: Warranty

    The scope of the servicing and the servicing intervals will depend on how often and to what degree the product is used. Service and maintenance must always be performed by Care of Sweden or one of its authorised technicians. Only use spare parts that have been approved by Care of Sweden. Use of non-approved spare parts invalidates the warranty.
  • Page 17: Other Information

    For more information contact Care of Sweden or your local dealer. 12.2 Returns and complaints Complaints are to be reported to the supplier. Return of a product shall follow the supplier’s instructions and can only be done after the complaint has been reported.
  • Page 18: Own Notes

    13. Own notes...