Product description ..... 12 Table of contents WARNINGS AND SAFETY Mattress and Control unit ..12 PRECAUTIONS ....... 1 Mattress........ 12 Introduction ......3 Control unit ......13 General information ....3 Air Cells and tubing set ..13 Intended purpose .....3 Reuse and cleaning ....
1. The product must only be installed and used for its intended purpose according to the instructions in this manual and/ or other documentation from Care of Sweden. The product must not be combined, assembled or repaired with parts (e.g. control unit and mattress), accessories or spare parts other than those described in this manual or other documentation from Care of Sweden.
Page 4
Contact an authorized service technician for examination and repair. 16. Never connect anything other than the Care of Sweden supplied Power supply to the control unit power cable connector.
2 Assembly and installation When unpacking, check that no parts are 3. Connect the air tubes damaged. If any damage is found, contact Care (marked ‘CPR’) coming of Sweden or your local distributor before using from the mattress on the the product.
3 Operation OPERATION CuroCell UNO® 1. Set the on/off switch on the side of the control unit to 1 (on). 2. Air will start to be pumped into the mattress. Depending on the size of the mattress it takes 20-40 minutes for the mattress to fully inflate.
pressure, the system returns to the previously selected program. At start-up, Pulsating Mode is always pre-set. Program There are three programs to choose from: Alternating mode means that the air pressure in the air cells are different and alternates regularly after chosen cycle periods. Pulsating mode combinates CLP with the The current selected settings are always alternating function.
Maximum pressure (caring mode) With this function, the entire mattress is inflated and provides fixed support. This function automatically switches to the preceding setting after approx. 20 minutes. This function should be used when nursing the user, shifting the user’s position, or moving the user in or out of bed.
Comfort setting CPR (cardiopulmonary The air pressure in the mattress can be adjusted in three stages: resuscitation) a basic setting set by the control In case of an emergency where CPR unit (0) according to the user’s (cardiopulmonary resus- weight, which can be increased in two stages citation) is necessary, (+1 or +2) according to the user’s wishes.
5. Carefully fold the mattress together, place • Information about the product the control unit between the folds of the • Contact information for Care of Sweden mattress and place in a transportation bag • Shortcut to quick tour (accessory) or equivalent for protective storage.
Major notifications cannot be acknowledged (disabled) without rectifying the error. The sound can be silenced temporarily via the function button but will return after approx. 5 min unless the error is rectified. In the event of a power failure and high system temperatures, information is not shown in the display, instead sound and light notify using the function button until the power is used up.
Page 13
Silent notifications Service If the control unit is in need of service, this is shown with an image. The service codes refer to different service actions, depending on which firmware version the control unit has installed. To find out which firmware version is installed, press the information button on the display. Always try to restart the device before contacting service.
4 Product description Mattress and Control unit 1. Mattress 2. CPR (Quick deflation) 3. Tube connection control unit 4. Control unit Mattress Hygienic cover, top part Air cell holder (integrated in inner cover) Air cells Safety mattress and tubing set Inner cover Cover, bottom part...
Control unit 1. Tube/CPR connection 4. Power switch, On/Off (1/0) 7. Air filter 2. Touch display 5. Connection power cable 8. Power adaptor 3. Function button 6. Hangers Air Cells and tubing set Static head air cells Press studs for inner cover Main air cells Heel air cells...
%, oxidizing solutions, chlorine max 1 % or reconditioning. Consult your hygiene manager hydrogen peroxide max 1,5 %. or Care of Sweden for help and instructions if you are unsure. MATTRESS 1. Disconnect the tube connector from...
7 Maintenance Service and maintenance must always be performed by Care of Sweden or one of its authorized technicians. Only use spare parts approved by Care of Sweden. For more 2. Remove the filter from the holder.
8 Troubleshooting If the problems keep occurring, please contact Care of Sweden or your local distributor. Problem Solution Check that the power adaptor has been connected to the mains The control unit does supply. not start Check that the LED on the power adaptor is showing green.
Symbol key Symbols to convey medical device information CE-marked in accordance with Medical Device Regulation (EU) Manufacturer 2017/745 Medical Device Distributor Symbols for traceability and product information Item number Typ BF SN-number IP class (enclosure class) Temperature range Symbols for user information Place directly on the bed base Foot placement Minimum length...
Page 21
Symbols for cleaning and recycling Wipe clean Machine wash 95°C Drip dry Chlorine Do not dry clean Do not iron Tumble dry Recycling Do not disope of with household waste...
A control unit with a battery backup must be disassembled and the battery removed for special battery recycling in accordance with local regulations. It is recommended that this is performed by authorised service technicians or that the control unit is returned to Care of Sweden for scrapping. Mattress A mattress should be sorted as combustible waste.