Page 1
Keysight UXG X-Series Agile Signal Generators Add Option AT2 (Attenuator) Kit Part Numbers: N5191-60501 - For Option 520 and 52E N5191-60502 - For Option 540 and 54E N5191-60503 - For Option 520 and 52E, 1EM N5191-60504 - For Option 540 and 54E, 1EM...
Page 2
The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
Page 3
Estimated Verification Time: ....Option 520/52E = 8 hours Option 540/54E = 15 hours Introduction This upgrade kit adds Option AT2 (Attenuator) to the UXG X-Series Agile Signal Generators. Installation includes the following major steps: 1. Check signal generator functionality, page 2.
Table 1 N5191-60501, for Option 520/52E Quantity Description Part Number UXG X-Series Agile Signal Generators Add AT2 to Option 520/52E this note and 540/54E Installation Note Output Leveling Assembly Option AT2/520 N5191-60370 Cable, RF, Semi-Rigid, Mod Filter Output to Coupler Input, Option...
Add Option AT2 (Attenuator) Upgrade Kit Check Signal Generator Functionality Use the following procedure to confirm that the signal generator powers up and the internal check identifies no errors. The internal check evaluates the operation of the signal generator and returns an error message if it detects a problem.
Page 7
Remove the Outer and Inner Bottom Rear Cover Remove the Outer and Inner Bottom Rear Cover Remove the Outer Cover Refer to Figure 1. Disconnect the power cord. 2. Using a T-20 driver, loosen the screws (1) and remove the strap handles (2) from both sides of the instrument.
Page 8
Remove the Outer and Inner Bottom Rear Cover Remove the Inner Bottom Rear Cover Refer to Figure 1. Place the signal generator upside down. 2. To remove from the rear bottom cover (1), use a T-10 driver to remove the twelve screws (2) (including two on the rear panel).
Page 9
Installing Option AT2 Installing Option AT2 Remove Existing A20 Output Leveling Assembly 1. For Option 520/52E instruments, refer to Figure For Option 540/54E instruments, refer to Figure For Option 520/52E 1EM instruments, refer to Figure For Option 540/54E 1EM instruments, refer to Figure 2.
Page 13
Installing Option AT2 Install A20 Output Leveling Assembly with AT2 Attenuator 1. For Option 520/52E instruments, refer to Figure For Option 540/54E instruments, refer to Figure For Option 520/52E 1EM instruments, refer to Figure For Option 540/54E 1EM instruments, refer to Figure 2.
Page 18
Activate Option AT2 Activate Option AT2 1. Turn the signal generator on and allow it to warm up for 5 minutes. 2. Follow the instructions on the Entitlement Certificate to redeem Option AT2. 3. Connect the instrument to the LAN and write down the IP address. 4.
Page 19
Verify Proper Installation of Option AT2 Verify Proper Installation of Option AT2 1. Press: System > Instrument Info > Options Info. 2. Make sure Option AT2 is listed. If you enable an option that does not have the required hardware installed, the menus for that option will be activated but the option cannot operate.
Page 20
Re-Assembling the Instrument Re-Assembling the Instrument Refer to Figure 1 Figure 1. Reinstall the inner bottom rear cover and outer instrument cover by reversing the order for removal. 2. Torque all T-10 screws to 9 in-lbs. 3. Torque all T-20 screws to 21 in-lbs. 4.
Verify Signal Generator Calibration Verify Signal Generator Calibration 1. Turn the signal generator on and allow it to warm up for 45 minutes. 2. Perform the following adjustment and performance tests, in the order listed, to verify signal generator calibration. Refer to the UXG Family Signal Generators Service Software. For Serial Prefixes <...