Basic Digital Operation (Option 653/655/656/657)
Understanding Option 012 (LO In/Out for Phase Coherency) with Multiple Baseband Generator
Synchronization
— The phase coherency feature only applies to the Dual ARB modulation mode.
— All cables from the splitter output to the instrument inputs should be of equal lengths.
Table 9-5
Option 012 (LO In/Out for Phase Coherency) Equipment
MIMO Configuration
2x2
3x3
4x4
All
a. On all of the MIMO configurations, the same length of SMA flexible cables are connected
from the splitter output to the inputs on the master and slave instruments. Refer to
9-30 on page 273
272
Cable Length
a
Part
n/a
As required
11636A
n/a
n/a
As required
PS3–20–
n/a
451/12S
n/a
As required
PS4–16–
n/a
452/10S
10502A
22.86 cm (9 inches)
and to
Figure 9-31 on page
Keysight EXG and MXG X-Series Signal Generators User's Guide
Notes
SMA flexible cables are connected from
the power splitter outputs to the LO
inputs on the rear panel of both the
master and the slave MXG/EXGs. Refer
to
Figure 9-30 on page 273
Power Divider, DC to 18 GHz. Refer to
www.keysight.com.
SMA flexible cables are connected from
the power splitter outputs to the LO
inputs on the rear panel of the slave
MXG/EXGs. Refer to
Figure 9-31 on
.
page 274
3–Way Pulser Microwave Corp., 3–Way
Wilkinson Dividers
The SMA flexible cables are connected
to the power splitter output to the LO
inputs on the rear panel of the slave
MXG/EXGs. Refer to
Figure 9-31 on
.
page 274
4–Way Pulser Microwave Corp., 4–Way
Wilkinson Dividers
Refer to
Figure 9-30 on page 273
Figure 9-31 on page 274
"Multiple Baseband Generator
Synchronization" on page 265
274.
.
and
. See also
.