ULM Troubleshooting

No communication with optical bench
Module will not fill
Module will not drain
Pump will not work
Module leaking

ULM Specifications

Item
Power
Temperature
Humidity
Dimensions
Weight
a. Visible portion of module once docked into the optical bench
b. Includes optical cell
PN B05577AC
Problem
• Module docked incorrectly. Eject module and re-dock. Apply
• Check to see if it is already full
• Check that connections are air-tight
• Check that the fill hose is submerged in the diluent.
• The valve should make an audible sound. If it does not,
• Check that the waste container is not full.
• Check that the drain hose is clear of obstructions.
• Module is not completely full of fluid.
• Check power to the system.
• Check belt drive.
• Check fittings.
• If leakage is inside, contact Beckman Coulter
• service.
The ULM derives its power directly from the optical bench < 50 W peak power
10° to 40° C (50° to 104° F)
0 to 90% without condensation
Height
49.5 cm (19.5 in)
8.85 kg (19.5 lb)
Possible Cause/Solution
slight pressure on the front of module while docking.
contact Beckman Coulter service.
Specification
Width
21.0 cm (8.25 in)
11.4 cm (4.5 in)
Sample Modules
Universal Liquid Module
a
b
Depth
Depth
6
6-85