Replacing the Diffraction Cell Window

Equipment Needed
• lens tissues
• lens cleaning solution
• window sealing tool
Procedure
Use this procedure to replace a diffraction sample cell window that is scratched or damaged.
To replace a diffraction sample cell window:
1
Drain liquid from module.
CAUTION
Wear appropriate safety equipment for the suspension fluids being handled.
2
Open the Optical Bench door and remove the sample cell.
3
Drain remaining liquid in the cell by disconnecting the two hoses from the line 1 and line 2 ports
on the sample cell front.
4
Loosen the two nylon screws to remove each window
PN B05577AC
Sample Modules
Aqueous Liquid Module
(Figure
6.10).
6
6-17