Sop), - Beckman Coulter LS 13 320 User Manual

Laser diffraction particle size analyzer
Table of Contents

Advertisement

Pump Speed
This option opens a dialog box that allows you to set the speed of the pump, by either clicking on
the arrows or by dragging the indicator.
Pump On – Pump Off
Selecting one of these options will turn the pump On or OFF.
PN B05577AC
Fill
This option opens the inlet valve to fill the vessel with the diluent.
The drain valve, if open, will be closed. Make sure the fill hose is
completely submerged in the diluent.
Rinse
Selecting this option will start an automatic rinse cycle. The order of
operations is Open Drain, Close Drain, Fill, and Pump On. This cycle
is repeated for a set number of times (set in the Run Cycle Options
dialog box).
De-bubble
This routine may be used to de-bubble the diluent when the
presence of too many bubbles produces high backgrounds.
6
Sample Modules
Universal Liquid Module
6-69