Performance Parameters And Features - NuAire NU-5510 Operation And Maintenance Manual

Air-jacketed automatic co2 incubator
Table of Contents

Advertisement

2.0

Performance Parameters and Features

Both the interior and exterior of the DHD Autoflow are constructed of 16-gauge material. The interior
is highly polished type 304 stainless steel, using crevice-free construction. All exposed edges are
deburred to insure no sharp edges. The exterior is cold rolled steel finished in a powder coated
polyurethane finish, which is resistant to chemicals and easily cleaned using mild household
detergents.
The DHD Autoflow's microcomputer temperature control system has two temperature sensors located
in the chamber. The temperature sensors compare the values to a set point and execute a differential
control algorithm that energizes a solid-state switch, supplying power to the heaters.
Up to 17 shelves can be placed inside the chamber (4 supplied).
Space-age high temperature material is used to insulate the inner chamber walls.
Foil heaters are directly attached to the chamber walls and the top and bottom surfaces of the inner
chamber.
Incubators can be stacked for space saving, and can still be serviced from the front.
Easily removable inner chamber plenum shelves and rails for sterilization.
Most electronics, motors, pumps, and valves are fully accessible from the front of the unit.
A CO
sample port is provided on the front panel to check the concentration of CO
2
The CO
percentage is controlled by a solid-state gas infrared sensor, which provides accurate
2
monitoring of CO
Automatic recovery of the CO
minute period.
The outer door includes an internal radiant heater in order to minimize condensation on the inner glass
door. A magnetic outer door gasket helps to insure a tight seal against the cabinet.
The inner glass door is 3/16" tempered with smooth-ground edges and seals are tight against a U-
grooved silicone rubber gasket. The door latch is cam action. An all solid-state magnetic switch
monitors door motion.
All control electronics are protected with a circuit breaker that may trip at 110% of loading rating but
will trip at 145% of load rating in less than 2 seconds. Should the circuit breaker open (pop-out button
will appear), merely depress to reset.
The Incubator has factory-installed adjustable leveling legs to compensate for uneven laboratory
surfaces.
OM0135
affected minimally by changes in temperature or humidity levels in the chamber.
2
level after a 1minute door opening to 5.0 + 0.2% is within a 4-1/2
2
5
in the chamber.