Starting The Cycle; Heat Up; Decontamination - NuAire NU-5510 Operation And Maintenance Manual

Air-jacketed automatic co2 incubator
Table of Contents

Advertisement

STARTING THE CYCLE

145
37
 Press the 95/145 button (for 3 seconds) until the Incubator displays show the following:
TEMP. display
CO2 display
The LED over the 95/145 button will flash on /off.
Indicates that user should:
Open the inner and outer doors
Empty water pan and put it back into the chamber.
Close the inner and the outer doors
Note: CO
system will be idle during the cycle. Protection of the sensor during the heated cycle is
2
automatic. The Decon Cycle will not advance to the heat up phase unless the door is opened to empty the
water pan and clean the chamber. When trying to advance the Incubator will beep rapidly and "dor" will
show in the Temperature display.

HEAT UP

 Press the 95/145 button again to start the heat up portion of the cycle.
Temperature display shows chamber temperature.
CO
2
The LED over the 95/145 button will flash on /off.

DECONTAMINATION

 The unit automatically starts a timer for the correct length of cycle when the decontamination
temperature is reached.
Temperature display shows chamber temperature.
CO
2
The LED over the 95/145 button is lit continuously.
OM0135
DRY 145 DEG C DECON CYCLE
TIME (HRS:MIN:SEC)
temperature continuous
H2O/out
3 hours
display flashes dCn
3 hours
display shows the time remaining for decontamination.
flashing
42