12 - Air Inject Cycle Time
This value specifies the amount of time in minutes
between each injection of air into the chamber
13 - Print Frequency Time
This parameter specifies the frequency, in minutes that lines
are to be printed on a status report. If the frequency is
specified as zero, no report will be printed.
14 - Temperature Time Out
This value determines the time, in minutes, for the temperature
to achieve set point. If the temperature doesn't get to within 0.2º
of set point within this time period, an alarm condition is declared.
15 - CO
Time Out
2
This value determines the time, in minutes, for the CO
percentage to achieve set point. If the CO
doesn't get to within 0.2% of set point within this time period,
an alarm condition is declared.
16 - Temperature Recovery Delay Time
This value determines the time, in seconds, to turn off the
main heater in half-degree increments during a temperature
recovery cycle. The delay time is required to prevent
temperature overshoot of the control set point.
17 - Temperature Sensor Differential
This value specifies a maximum differential, measured in
temperature (C) that the two temperature sensors may deviate
from one another, or from the last read value. If this
differential is exceeded, a warning LED is shown on the
running chamber. If multiple sensors fail to read within the
specified limits, an alarm condition is declared. An alarm
condition will cause the chamber to enter a safe condition
where no power is enabled to any of the system output controls
until the situation is rectified.
OM0135