Air Inlet Connection; Co2 Connection; Co2 Tube Connection; Co2 Supply - NuAire NU-5510 Operation And Maintenance Manual

Air-jacketed automatic co2 incubator
Table of Contents

Advertisement

7.5

Air Inlet Connection

An air inlet tubing kit consists of one-foot clear vinyl tube and (1) 50mm polypropylene .3-micron HEPA
filter. Locate the air inlet port on the back panel of the unit. Remove the cover cap; connect one end of
tubing to the air inlet port, and the other end of tubing to the air filter.
CAUTION: This is a free air supply. DO NOT connect to pressurized source.
7.6
CO
Connection
2
High concentrations of CO
This Incubator is designed to be operated with CO
result in a hazardous condition. Gases other than CO
cylinders have a UN1013 label on the cylinder and are equipped with a CGA 320 outlet valve. Check the
gas cylinder for the proper identification labels.
Do not use CO
gas cylinders equipped with siphon tubes. A siphon tube is used to extract liquid CO
2
the cylinder which can damage the pressure regulator. Consult with your gas supplier to ensure that the
CO
cylinder does no contain a siphon tube.
2
7.6.1
CO
Tube Connection
2
Included with the Incubator are a tubing kit consisting of (1) six foot (2m) vinyl tube and
(1) 50mm polypropylene 0.3 micron HEPA filter.
7.6.2
CO
Supply
2
1. CO
2. A two-stage pressure regulator, Linde# 19590, or equal, is recommended.
3. DO NOT USE a single stage regulator. It will not give a stable output at 20 psi and
OM0135
gas can cause asphyxiation! Install Incubator in well ventilated area.
2
of medical grade is recommended.
2
exposes the Incubator to the gas cylinder pressure.
gas only. Connecting a flammable or toxic gas can
2
should not be connected to this equipment. CO
2
14
gas
2
from
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nu-5510e

Table of Contents