Humidified 95 Deg. C Decon Cycle - NuAire NU-5510 Operation And Maintenance Manual

Air-jacketed automatic co2 incubator
Table of Contents

Advertisement

HUMIDIFIED 95 DEG. C DECON CYCLE

95
37
STARTING THE CYCLE
 Press the 95/145 button (for 3 seconds) until the Incubator displays show the following:
TEMP. display
CO
2
The LED over the 95/145 button will flash on /off.
Indicates that user should:
Open the inner and outer doors
Empty water pan disinfect it and refill with 300 ml of distilled water. Place it back in the
chamber on the highest shelf.
Close the inner and the outer doors
Note: CO
system will be idle during the cycle. Protection of the sensor during the heated cycle is
2
automatic. The Decon Cycle will not advance to the heat up phase unless the door is opened to empty the
water pan and clean the chamber. When trying to advance the Incubator will beep rapidly and "dor" will
show in the Temperature display.
OM0135
TEMPERATURE
RELATIVE HUMIDITY
Decon Phase
TIME (HR:MN:SEC)
H2O
display
out
flashing
flashing
45
Drying Phase