2 - CO
Tank 2 Enable (Optional)
2
This function will enable or disable the optional CO
tank 2 system. The value display will show "on" or "off"
corresponding to the current condition.
3 - CO
Tank Switch Back (option)
2
Note: function can only be enabled with the CO
option in use. Unit will automatically check tank 1 for
gas pressure every 12 hours, and stay on tank 1 if
pressure is detected
4 - Closed Door CO
Zero/Span Calibration
2
This option enables user to run zero & span calibration
on the CO
sensor with out opening the Incubator door.
2
(See section 9.4.2).
5 - Alarm Audible Enable
This function will enable or disable the audible alarm
ring back function. The value display will show "ON"
or "OFF" corresponding to the current condition. If the
function is "ON"; the audible alarm will provide a ring back
of the alarm condition. If the user pushes any key to silence
the audible alarm, after 15 minutes of silence the audible
alarm will return. If the function is "OFF", the ring back of
the alarm condition will never come back after the user
pushes a key to silence the audible alarm.
6 - Decon Cycle selection
This function allows the user to select either the 145 dry
or 95 moist Decon Cycle by pushing the arrows.
Follow all instructions in section 10.1 of the manual
for either of the cycles chosen.
OM0135
2
tank 2
2
30