18 - Temperature Maximum Above Set point

This value determines the maximum deviation, measured
in (C) that the chamber is permitted to rise above set point,
once it is researched before an alarm condition is declared.
An alarm condition will cause the chamber to enter a safe
condition where no power is enabled to any of the system
output controls until the situation is rectified.
19 - CO

Maximum Above Set point

2
This value determines the maximum deviation, measured in
CO
percent (%) that the chamber is permitted above or below,
2
once the Incubator reaches the specified set point before an
alarm condition is declared.
8.8.1.3
OM0135

Reset, Master

The master reset diagnostic function is the last effort to correct operational faults which
otherwise cannot be solved.
By reloading the default configuration, the entire memory will be reset and ALL
CALIBRATION OFFSETS AND CONFIGURATION OPTIONS WILL BE LOST.
All calibrations will need to be performed following a master reset.
To reset: At master reset menu, hit [SEL] key twice, turn off Incubator for about 10
seconds, and then turn it back on to complete master reset.