Door Heater; Perimeter Heater; Air Inject Valve; Air Pump Relay - NuAire NU-5510 Operation And Maintenance Manual

Air-jacketed automatic co2 incubator
Table of Contents

Advertisement

7 - CO
Tank 2 Valve
2
This function shows the current state of the CO
The CO
percent display will show "on" or "off" corresponding
2
to the valve condition.

8 - Door Heater

This function shows the current state of the door heater.
This function also allows the door heater to be turned on at
different percentages (0, 25, 50, 75, 100) alternating with
the "dor" indicator.

9 - Perimeter Heater

This function shows the current state of the perimeter heater. This
function also allows the perimeter heater to be turned on at different
percentages (0, 25, 50, 75, 100) alternating with "the PEr" indicator.

10 - Air Inject Valve

This function shows the current state of the air inject valve.
The CO
percent display will show "on" or "off" corresponding
2
to the relay condition.

11- Air Pump Relay

This function shows the current state of the relay that turns the
air pump on and off.
12 - Alarm Relay
This function shows the current state of the alarm relay.
The CO
percent display will show "on" or "off" corresponding
2
to the relay condition.
13 - Sensor Shroud Temperature
This function shows the temperature of CO
the temperature display. The CO
the heater relay as being on or off. The temp displayed should be
at least chamber temperature + 1 deg. on an Incubator in stable condition.
It can display as high as 57 deg. C.
OM0135
tank 2 valve.
2
sensor shroud in
2
display shows the state of
2
28