BER Testing

3 Connect the other end(s) of the adaptor cable to the test access point.
You have finished connecting the cables.
Performing a bit error rate test (BERT) involves configuring the test settings,
including logic error insertion; connecting the E1/Datacom Tester to the a
circuit; starting the test and inserting logic errors; and observing the results.
To do a bit error rate test
1 Set the test mode to DTE or DCE Emulation. See
mode" on page
53.
2 Select a pattern. See
3 Specify the other settings on the Configuration menu, as needed.
See
"Specifying the interface standard" on page
See
"Specifying timing parameters" on page
See
"Specifying polarity settings" on page
See
"Specifying data parameters" on page
See
"Specifying async settings" on page
See
"Specifying flow control settings" on page
See
"Specifying G.821 performance settings" on page 60
4 Specify how errors will be inserted. See
page
60.
5 If you want to run the test for a specific amount of time, configure the timed
test settings. See
"Timed testing" on page
6 If this is your first test of the day, perform a self test on the E1/Datacom
Tester. See
"Performing a self test" on page
7 Connect the E1/Datacom Tester to the circuit. See
Datacom Tester to the circuit" on page
8 From the Action menu, select Restart.
9 Verify that the Sync LED illuminates green.
10 To insert errors, press the * Action key to view the Error menu, and then
select press 1.
This will do one of the following:
– Enable or Disable Bit Error Rate Insertion — begin inserting bit
errors at the rate specified on the Error Settings menu (appears only if
the Rate Type is set to "Rate").
– Bit Error Multi Insert — inserts multiple bit errors with each press
(appears only if the Rate Type is set to "Multiple").
– Bit Error Insert — inserts one bit error with each press (appears only if
the Rate Type is set to "Single").
11 From the BERT menu, select Results and then the BERT category.
SmartClass E1 Tester User's Guide
"Specifying a BERT pattern" on page
54.
55.
57.
58.
"Specifying error settings" on
25.
61.
61.
Chapter 5 Datacom Testing

BER Testing

"Specifying the test
55.
53.
59.
"Connecting the E1/
63