This probe parameter refers to the resistance of the control probe at 0°C. Nor-
mally this is set for 100.000 ohms.

ALPHA

This probe parameter refers to the average sensitivity of the probe between 0
and 100°C. Normally this is set for 0.00385°C

Operating parameters

The operating parameters menu is indicated by,
Operating parameters menu
PAr
Press "SET" to enter the menu. The operating parameters menu contains the
cut-out reset mode parameter.

Cut-out reset mode

The cut-out reset mode determines whether the cut-out resets automatically
when the bath temperature drops to a safe value or must be manually reset by
the operator.
The parameter is indicated by,
Cut-out reset mode parameter
CtorSt
Press "SET" to access the parameter setting. Normally the cut-out is set for
manual mode.
Cut-out set for manual reset
Cto=rst
To change to automatic reset mode press "UP" and then "SET".
Cut-out set for automatic reset
Cto=Auto

Stir mode select

This parameter along with the Stirrer Set-point allows the user to set the tem-
perature at which the stirrer motor is activated. This setting is generally used
when salt is used for the bath medium. For example, you can set the mode to
"auto" and the temperature to 200°C. This allows the stirrer motor to shut off
and turn on only when the salt is a liquid (>200°C) preventing the stirrer motor
from overheating and damage.
The parameter is indicated by,
-1
.