4. Remove the module cover.
Loosen the captive screw using a #2 Phillips screwdriver to remove the module
cover.
Note
5. Remove the module you want to replace.
Slide the module out of the housing by inserting a finger in the slot at each end of
the module and gently pull the module away from the housing until you completely
remove it.
6. Disconnect the USB cable from the rear of the module.

Reassembly notes

When aligning the module with the housing, the USB connection is on the left.
The NIBP module must always be in the front-most slot.
The modules fit tightly in the housing but can become loose
over time. Support the modules to prevent them from falling
out as you remove the module cover.