Functional Verification And Calibration - Welch Allyn 6000 Series Service Manual

Vital signs monitor
Hide thumbs Also See for 6000 Series:
Table of Contents

Advertisement

Functional verification tests
The functional verification tests help to confirm the proper operation of the monitor and its options.
These tests support the requirements of routine preventive maintenance. It is not necessary to
disassemble the monitor to perform these tests.
The tests may also be useful as a diagnostic tool to help isolate a malfunction.
Each time you open the case and prior to returning the monitor to service, use the service tool,
Gold edition, to complete the full suite of tests to ensure that all systems are operating within the
design specifications.

Functional verification and calibration

Required equipment
Material no.
01802-110
DOC-10
SRC-MAX
06138-000
4500-30
5082-183
103507
4500-925
407691
541-USB001
407022
713549
Description
Tester, calibration, 9600 plus
Cable, SP02 extension, Nellcor
Nellcor portable oximetry tester MP506
CAL-KEY, assembly, M690/692
Blood pressure hose, 5 ft.
BP Y-tube, no fittings 1/8 in tube
Masimo rainbow tester (Masimo part #2368)
USB 2.0/5-pin type A to mini-B cable, gold, 6 feet
Service test box
OAE, USB 6' cable
Assy, MMF Nurse call cable - Service
USB 2.0 Dual A to Single Mini B Cable
81
Qty
3
1
1
1
1
1
1
1
1
1
1
1

Hide quick links: