1
Installation
Table 6
Power consumption
Product
Line voltage (VAC)
111 Vial model
100*/120
single phase 
(–10% / +10%)
111 Vial model
200/220/230/240
single/split phase
(–10% / +10%)
12 Vial model
100*/120
single phase 
(–10% / +10%)
12 Vial model
200/220/230/240
single/split phase
(–10% / +10%)
* Users of 100 V power (Japan) receive the 120 V US power option with a 100 V/120 V step up transformer.

Install the configuration plug

WA R N I N G
C A U T I O N
18
Frequency (Hz)
Maximum
continuous
power
consumption
(VA)
50/60
850
50/60
850
50/60
850
50/60
850
Install the configuration plug as follows:
Only an Agilent-trained technician should perform this procedure.
Components can be damaged by static electricity: be sure to wear
an ESD strap grounded to the GC chassis while performing this
procedure.
Current rating
Power outlet
(amps)
current rating
6.2
15 Amp
3.8/3.4/3.3/3.1
10 Amp
6.2
15 Amp
3.8/3.4/3.3/3.1
10 Amp
Installation and First Startup