Using a Swagelok Tee

N O T E
Installation and First Startup
To supply gas from a single source to more than one input,
use a Swagelok Tee.
Do not combine valve actuator air with flame ionization air. The valve
action will cause major upsets in the detector signal.
Materials needed:
• 1/8- inch preconditioned copper tubing
• Tubing cutter
• 1/8- inch Swagelok nuts and front and back ferrules
• 1/8- inch Swagelok Tee
• Two 7/16- inch wrenches
• 1/8- inch Swagelok cap (optional)
1
Cut the tubing where you want to install the Tee. Connect
the tubing and Tee with a Swagelok fitting. See
Cap
Tee fitting
Nut and
ferrules
Figure 31 Swagelok tee
Measure the distance from the Tee to the instrument
2
fittings. Attach copper tubing to the open Tee ends with
Swagelok fittings.
Making Swagelok Connections
Figure
A
31.
73