Configure the Headspace Sampler
Installation and First Startup
Next, enter details that describe the gas types, vial sizes,
sample loop size, and other important information.
Press [Config].
1
Use
to scroll to Vial Gas type and press [Mode/Type].
2
Use
to select the vial pressurization gas type from the
3
list, then press [Enter].
Scroll to Loop Volume (mL). Press [Mode/Type]. Select 1.0 as
4
the sample loop volume, in mL, then press [Enter]. (All
headspace samplers ship with a 1 mL sample loop
installed.)
If you ordered the optional G4562A Carrier Gas EPC
5
Module Accessory, scroll to Carrier. Press [Mode/Type].
Select the correct control mode for checkout:
•
If the HS will provide all carrier gas to the GC, select
GC Control. (For example, you did not plumb carrier gas
to the GC.)
•
If the GC will provide its own carrier gas flow, select
GC + HS Control. In this case, the HS will provide enough
additional carrier gas flow during injection to push the
checkout sample into the GC inlet.
Scroll to APG polarity. If needed, press [Mode/Type] to
6
change.
•
For Agilent GCs, use Active high.
•
For non- Agilent GCs, refer to the GC documentation.
1
Installation
51