Connect Gas Supplies

N O T E
1/8-in. flare
Vial
Carrier gas
pressurization
supply
gas supply
(N2, He, H2,
(N2, He)
Ar/Me)
Swagelok cap at inlet
30
Agilent recommends installing shutoff valves (not supplied) near the vial
pressurization and carrier gas fittings. Using the valves provides a fast way
to isolate the HS during troubleshooting and periods of inactivity.
Agilent provides hardware to plumb the instrument gases as
shown in
Figure
10. In this example, the HS connects to a
7890A GC through the inlet septum. The GC provides carrier
gas control. The vial pressurization gas is connected to a
different supply than the carrier gas.
Carrier gas from
GC inlet
1/16-inch union
0100-0124
Figure 10 Standard plumbing configuration
See
Table 8
for a summary of the acceptable gas types for
the HS.
1530-2239 Tubing, post
drawn, 1650 mm
1530-2239 Tubing, post
drawn, 1650 mm
1/8-in. flare