Run The Checkout Sample - Agilent Technologies 7697A Installation And Startup Manual

Headspace sampler
Hide thumbs Also See for 7697A:
Table of Contents

Advertisement

2
First Startup

Run the Checkout Sample

1
62
After the HS has been Ready for 25–30 minutes, it has been
purged and is sufficiently stable to run the checkout sample.
Load the checkout sample into the HS.
1
For the 111 vial model:
Press [Tray Park/Carousel Advance] to "park" the tray
a
(move the gantry to a rest position for easy access to
the vial racks).
Place the capped sample vial in position 1.
b
Figure 19 Tray vial position 1
Press [Tray Park/Carousel Advance] to prepare the tray for
c
use.
For the 12 vial model:
Open the tray cover. A label to the left of each vial
a
position denotes the number of that position. See
Figure 19
and
If vial position 1 is not available, press [Tray
b
Park/Carousel Advance] to rotate the tray.
Place the capped sample vial in position 1.
c
Figure
20.
Installation and First Startup
1