Attach the transfer line to a GC inlet

C A U T I O N
WA R N I N G
Installation and First Startup
Wear clean, lint-free gloves to prevent contamination of parts with
dirt and skin oils.
Be careful! The oven and/or inlet may be hot enough to cause
burns. If either is hot, wear heat-resistant gloves to protect your
hands.
Split Splitless or Multimode Inlet
Slide the one- piece septum nut and strain relief coupling
1
onto the tubing until the inner metal sleeve of the
transfer line bottoms- out on the strain relief.
Slide the septum onto the tubing until the septum is
2
against the bottom of the one- piece septum nut and
strain relief coupling.
Trim the tubing 35 mm past the end of the septum.
3
Install the inlet liner (with o- ring) that is appropriate for
4
your application.
5
Insert the tip of the transfer line into the inlet.
6
Finger- tighten the strain relief coupling.
Purged Packed Inlet
Slide the one- piece septum nut and strain relief coupling
1
onto the tubing until the inner metal sleeve of the
transfer line bottoms- out on the strain relief.
Slide the septum onto the tubing until the septum is
2
against the bottom of the one- piece septum nut and
strain relief coupling.
Trim the tubing 2 mm past the end of the septum.
3
Install the inlet liner (with o- ring) that is appropriate for
4
your application.
Insert the tip of the transfer line into the inlet.
5
Finger- tighten the strain relief coupling.
6