Install The Configuration Plug; Verify The Available Power Configuration - Agilent Technologies 7697A Installation And Startup Manual

Headspace sampler
Hide thumbs Also See for 7697A:
Table of Contents

Advertisement

Install the Configuration Plug

Verify the available power configuration

WA R N I N G
Installation and First Startup
Before use, set the HS power configuration.
Locate the configuration plug that shipped with the
1
headspace sampler. Compare the voltage shown on the
configuration plug label with the laboratory line voltage.
(For customers in Japan using 100 VAC, the unit is
120 VAC with an external 100 V/200 V step up
transformer.) See
Table 5
Headspace sampler power configuration plugs
Part number
Voltage
G4556-60390
120 VAC
G4556-60400
200 VAC
G4556-60410
220 VAC
G4556-60420
230 VAC
G4556-60430
240 VAC
Locate the power cord that shipped with the headspace
2
sampler. Verify the power cord is correct for the locale
and voltage.
Measuring the mains voltage can expose you to hazardous
electrical current. Only a trained technician should perform this
step.
Measure the power outlet voltage and verify it meets the
tolerance requirements listed in
Table
5.
Color
Brown
Red
Orange
Yellow
Violet
Table
6.
1
Installation
17