First Startup

Tools and Materials Required

54
Materials needed:
• Headspace checkout sample, 5182- 9733
• Flat bottom headspace vial, 20 mL
• Crimp cap, headspace, 20 mm septum
Tools required:
• Ergonomic manual crimper for 20 mm caps
• Forceps or tweezers
The checkout sample kit, part number 5182- 9733, contains
5 mL of sample and a box of 5- µL micropipettes.
The sample composition is:
• 9.59 ± 0.05% n- nonane (w/w)
• 0.348 ± 0.002% 1,2- dichlorobenzene (w/w)
• 0.322 ± 0.002% nitrobenzene (w/w)
• 0.241 ± 0.001% tert- butyldisulfide (w/w)
in dodecane matrix.