Usage Of The As-Probe Adapter - Projekt Elektronik FM 302 Operating Manual

Teslameter
Table of Contents

Advertisement

6. Operation
wire color
white
yellow
green
brown
One should be aware, that the probes and all outputs have a common
ground. Especially when using the brass version of AS-probes (AS-
NTM, AS-LTM) an isolation between probe and other parts of the
measurement setup can be necessary.
6.5.4

Usage of the AS-probe adapter

To use the AS-probe adapter three connections have to be made.
At first the AS-active-probe is connected via the delivered adapter cable with
the connectors "±3V" and "IN" of the AS-probe adapter.
For connection assignment of the adapter cable see Fehler! Verweisquelle
konnte nicht gefunden werden. on page Fehler! Textmarke nicht
definiert..
For power supply the input "PWR" has to be connected to a power supply
which delivers a DC voltage between 9 V and 36 V.
As third the output "OUT" has to be connected with the analog input of a
data acquisition system like e.g. a PLC.
The gain switch is set to the desired position (x5 or x50).
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
probe function
probe signal
+3 V
-3 V
ground
Table 1
Figure 61 Connection AS-probe adapter
connector adapter
IN ±
±3V +
±3V -
IN ┴
Page 75

Advertisement

Table of Contents
loading

Table of Contents