Transverse Probe Hot With Improved Temperature Characteristics 2000 Mt (As-Ntp-Hot-05) - Projekt Elektronik FM 302 Operating Manual

Teslameter
Table of Contents

Advertisement

7. Technical Specifications
7.2.12 Transverse Probe Hot with Improved Temperature
Characteristics 2000 mT (AS-NTP-Hot-05)
Figure 81 Size transverse probe Hot 2000 mT (AS-NTP-Hot-05)
Ranges (with FM 302)
Effective area
Thickness of support
Length of support
Width of support
Transfer factor
Bandwidth (- 3 dB)
Rise time
Linearity error
Temperature error
Noise
Operation temperature
Storage temperature
Max. relative humidity
Power
Connector
Output impedance
Length of cable
Technical specifications are subject to change without prior notice!
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
±20 mT; ±200 mT; ±2000 mT
2
0.5 mm
1.5 ±0.1 mm
70 mm
5 ±0.5 mm
1 V/T
0 - 35 kHz
<3 µs
<0.2 % ±0.2 mT (at 20 °C ± 1 °C)
<±1.0 % ±0.2 mT (-10 °C to +150 °C)
typ. 21 µT
(10 Hz – 10 kHz)
RMS
typ. 18 µT
(DC – 10 Hz, 50 s)
PP
-40 °C to +150 °C (only probe, grip and cable)
+5 °C to +50 °C (probe connector)
-10 °C to +60 °C
70 % at +35 °C
±3 V through FM 302, AS-probe adapter,
AS-Adapter 3 or PLC
15 pol. SubD
<1 Ω
2.95 m