Geo-X Axial Probe 200 Μt (As-Uap Geo-X) - Projekt Elektronik FM 302 Operating Manual

Teslameter
Table of Contents

Advertisement

7. Technical Specifications
7.2.18 GEO-X Axial Probe 200 µT (AS-UAP GEO-X)
Ranges (with FM 302)
Effective volume
Case diameter
Case length
Transfer factor
Bandwidth (-3 dB)
Rise time
Linearity error
Temperature coefficient
Zero drift
Hysteresis
magnetic flux density
Noise
Operation temperature
Storage temperature
Max. relative humidity
Power
Connector
Output impedance
Length of cable
Technical specifications are subject to change without prior notice!
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Figure 87 Size GEO-X axial probe 200 µT (AS-UAP GEO-X)
±2 µT; ±20 µT; ±200 µT
Ø 5 mm x 22 mm
Ø 17 mm
64 mm
1 V / 100 µT
0 – 800 Hz (2 µT- / 20 µT-range)
0 – 500 Hz (200 µT-range)
<0.3 ms
<0.8 % ±0.2 µT (at 25 °C)
max. ±0.1 %/K (10 °C to 50 °C)
max. ±10 nT/K
max. 0.1 % of value
max. ±200 µT or 140 µT
typ. 4,5 nT
(10 Hz – 1 kHz)
RMS
typ. 6 nT
(DC – 10 Hz, 50 s)
PP
+5 °C to +50 °C
-10 °C to +60 °C
70 % at +35 °C
±3 V through FM 302, AS-probe adapter,
AS-Adapter 3 or PLC
15 pol. SubD
<1 Ω
1.5 m
eff