6. Operation

6.4.12 Usage of the AS-Active-Probes

6.4.12.1 Usage with the Teslameter FM 302

Usually the AS-active-probe is simply connected to the Teslameter. The
measurement can be started immediately.
Also all extended possibilities of the Teslameter FM 302 like calibrated analog
output, control via USB interface or power supply with power adapter are
usable in that way. Further details can be found in section 6.3.2 Usage of The
Teslameter FM 302.

6.4.12.2 Usage as Autonomous Transducer

Our AS-active-probes can be operated stand-alone. Therefore they simply
have to be supplied with ± 3 V (± 1 %) with max. 20 mA. The analog output
signal can be feed e.g. into the input of a programmable amplifier of a PLC.
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Figure 52 Usage of AS-active-probe with FM 302