Command "Version; Command "Zero - Projekt Elektronik FM 302 Operating Manual

Teslameter
Table of Contents

Advertisement

6. Operation

6.3.3.23 Command "version"

command:
without
parameter
example:
version
firmware version is V1.0, built Sep

6.3.3.24 Command "zero"

command:
without
parameter
with parameter
parameter
see also section 6.3.1.7 Key "zero" – Offset Compensation
examples:
zero
zero compensation value is -7365
zero set
zeroing.....
zero compensation value is 10610
zero -12345
zero compensation value is -12345
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
version
shows the version of the installed firmware
zero [{offset|set}]
shows the current value of the offset compensation
With
"set"
as
compensation process is started.
If a number is given as parameter, it is taken as new
value for the offset compensation. The exact transfer
factor is device dependent.
set, -39320 ≤ offset ≤ 26213
8 2011, 13:10:34
parameter
the
automatic
offset
Page 48