6. Operation

6.4.12.3 Usage with the AS-Probe Adapter

To simplify the usage of the AS-active-probe as autonomous transducer, the
AS-probe adapter can be used. It provides high stable ±3 V to supply the
probe and amplifies the output signal to ±10 V.
For further details see section 6.6.4 Usage of the AS-Adapter.

6.4.12.4 Usage with the AS-Adapter 3

Up to three AS probes simultaneously can be operated on the AS-Adapter
3. The AS-Adapter 3 ensures the supply of all connected probes.
Further details can be found in chapter 6.6.4 Using the AS-Adapter 3

6.4.13 Zero Chamber (optional)

Optional a zero chamber is attainable for our instruments.
The zero chamber is a one side closed pipe of good magnetic shielding
metal to shield the existing outer field. That is at least the earth magnetic
field. In addition there may be other interfering fields from the environment.
In real world shielding may not be 100 %. A small residual magnetic field
remains inside of the zero chamber.
If necessary the zero point of an AS-active-probe may be checked and
adjusted with the help of a zero chamber. Therefor the AS-probe is feed into
the zero chamber. Now one can assume that the magnetic field is sufficient
shielded. With the "zero" key of the Teslameter FM 302 (see section 6.3.1.7
Key "zero" – Offset Compensation) the display may set to zero in the most
sensitive range.

6.4.14 Linearity Curves (optional)

Linearity curves are optionally available for the AS-active-probes. The
linearity curves are used for determining exact field values at up to five
Projekt Elektronik
MESS - UND REGELUNGSTECHNIK GMBH
Figure 55 Structure AS-active-probe
Figure 56 Zero Chamber
Page 69