3.5

SB Series

Follow the steps to set up and calibrate the Ishida SB Series scale.
Step
1
Turn on scale.
2
Press [CAL] , then release the key.
3
Press and hold the T key, then press the F key, then release both keys.
4
The scale will automatically be set to the zero point.
5
After the zeroing procedure is finished, the measurement point must be set.
6
Set standard weight in center of weigh pan. The scale will automatically calibrate
itself.
7
After calibration has finished, scale will automatically return to the original
measurement mode.
NOTES:
The Ishida SB Series precision balance automatically calibrates itself.
Press the
P
key to cancel calibration procedures and return to the weigh display.
If
[o-Err]
appears on the display, the weight exceeds the scale capacity.
Standard calibration weight is
If
appears on the display, the calibration weight is less than 50% of the standard.
[1-Err]
If
[2-Err
] appears on the display, an object other than the calibration weight is on the pan, or the display
tolerance (1%) is exceeded.
20
Ishida Setup & Calibration Manual
Action
capacity. However, when possible, calibrate using full capacity.
1/2
Display
00000 00000 00000
CAL
on 0
on F.S
on F.S
5000.00