Follow the steps below to set up and calibrate the NOVA-II scale.
IMPORTANT!
All work performed on the NOVA-II, including setup and calibration, must be done by an authorized
scale distributor to ensure compliance with NTEP requirements.
Step
1
Turn the scale off.
2
Turn the scale on and immediately press the 1 key.
3
Press the 1 key then the TARE key.
4
Press TARE after verifying the country code: 20 = United States (15 lb, 30 lb, and 60 lb)
To change the country code, enter 20 20, then press TARE.
5
Press TARE after verifying the Capacity Code:
• 2 = 30 x 0.01 lb
• 9 = 15 x 0.005 lb
• 7 = 60 x 0.02 lb
NOTE: XXXX = capacity, z = code. To change, enter the code, then press TARE.
6
With no weight on the scale, press ZERO to set 2000 counts. NOTE: nnnn = any number.
7
Place the full capacity weight on the scale as determined in Step 5.
8
Press TARE. NOTE: nnnn = any number
9
The center display should read 32000 ± 1. If not, repeat the above starting with Step 6.
10
Remove the weights and the weigh platter.
11
Tip the scale on edge and remove the brass screw. See the diagram below. NOTE: nnn =
any number.
12
Use a non-metallic device to press the memory switch on the circuit board. See the
diagram below.
The scale beeps and temporarily displays "ok."
13
Replace the brass screw and the weigh platter. Turn off the scale.
14
Turn the scale on again and test the calibration by placing a known weight on the platter.
2
Ishida Setup & Calibration Manual
Action
Memory Switch
Figure 1-1. Bottom view of NOVA-II scale
Display
C
0
C1-1 20 0
C1-2 XX.XX
z
10 nnnnn nnnn
10 2000 0
10 nnnnn nnnn
10 32000
30000
10 32000
30000
10 nnnnn nnnn
10 nnnnn nnnn
10
s
ok
10
2000
0
0.00 0.00 0.00