Ishida distributor site at intended for use by service technicians responsible for www.ishidaretail.com installing and servicing these products. Retail Scales Rice Lake Weighing Systems currently sells and distributes the following models of Ishida retail scales. • Nova–II • BC–3000 The following pages describe the setup and calibration procedures used for these products.
NOVA-II Follow the steps below to set up and calibrate the NOVA-II scale. IMPORTANT! All work performed on the NOVA-II, including setup and calibration, must be done by an authorized scale distributor to ensure compliance with NTEP requirements. Step Action Display Turn the scale off.
BC-3000 Follow the steps below to set up and calibrate the Ishida BC-3000 scale. Step Action Display Turn off scale. Turn on scale while depressing the 1 key. Release the 1 key after two beeps. Press the 1 key and push ENTER two times. xxxx=default values between 500 and 5000.
Industrial Scales Rice Lake Weighing Systems sells and distributes the following models of Ishida industrial scales. • MTX Series • MTB Series • MTE Series • MG-3000 Series • CWS Series • NCS Series • QBX Series The following pages describe the setup and calibration procedures for these products.
4. If the C8–C5 settings are not correct, turn the scale off, then on again and re-enter the settings. If the C8–C5 settings are correct, press the key. COUNT Calibration Continue with normal calibration as described on page7, Section 2.1.2 beginning from Step 3 (display = CA–). Ishida Setup & Calibration Manual...
Page 11
Follow the steps in the table below to calibrate the MTX/MTB Series scales. NOTE: Configuration settings must be set prior to calibrating the MTX/MTB series scale. 2.1.2 Calibration Step Action Display Press and hold the ZERO and TARE keys, then press the ON/OFF key. Release the ON/OFF key first, then release both the ZERO and TARE keys.
When you mount the display pole on the main body, secure the load cell cable with the plastic tie as shown below. After you secure the cable, use the procedures (shown above) to calibrate the MTE Series scale. Plastic Tie Load Cell Cable Figure 2-2. Bottom view of MTE Series scale base Ishida Setup & Calibration Manual...
MG-3000 The MG-3000 uses one of two CPU boards. Early versions used the circuit board number P-780; later versions use the circuit board number P-802. Both function the same, but the calibration and setup procedures differ. 2.3.1 Calibration For Units With P-802 Circuit Board Follow the steps below to calibrate an MG-3000 with a P–802 circuit board.
Page 14
The switch feels like it clicks when pressed. Turn on the scale and test the calibration by placing a known weight on the scale. 17000 When calibration is complete, replace the brass hex bolt on the back of the indicator. Ishida Setup & Calibration Manual...
CWS Series Follow the steps below to set up and calibrate the CWS Series scale. KG and GRAM Calibration Procedure: Step Action Display Turn off the scale. Press and hold the TARE and ZERO keys, then press and release the ON/OFF key. Release the TARE and ZERO after the display goes blank.
Turn the scale on again and test the calibration by placing a known weight on the scale. Memory Switch CWS-oz-05 CWS-oz-10 CWS-oz-19 Capacity Code Capacity Code Capacity Code 5 Lb x 0.1 10 lb x 0.2 oz 19 lb x 0.5 oz Table 2-2. Lb/Oz Scale capacity codes Ishida Setup & Calibration Manual...
NCS Series Follow the steps below to set up and calibrate the CWS Series scale. Step Action Display Turn the scale off. Turn the scale on and immediately press the 1 key. Press the 1 key, then the TARE key. Press TARE after verifying the country code: 20 = United States (6 lb, 30 lb, and 60 lb).
Balances Rice Lake Weighing Systems sells and distributes the following models of Ishida precision balances. • QB Series • MB Series • IB Series • SB Series The following pages describe the setup and calibration procedures for these products. Balances...
50% of the standard. • [2-Err] appears on the display, the display tolerance (0.4%) is exceeded, or the scale may be defective. Contact your Ishida representative. Unit No. (mark) MB-150E MB-300E...
, this indicates that the error at span [4-Err] [WEIGH] varies too much from that of the factory-set value. Unit No. (mark) MB-300RE MB-600RE 1 (g) 2 (ct) 1500 3000 3 (Lb/oz) Figure 3-1. Calibration Knob Ishida Setup & Calibration Manual...
IB Series Follow the steps below to set up and calibrate the Ishida IB Series scale. CALIBRATION Step Action Display Turn on scale. 0000 0000 0000 Press and hold the F key until the [CAL] message is displayed. Press and hold the T key, press the F key, then release both keys.
SB Series Follow the steps to set up and calibrate the Ishida SB Series scale. Step Action Display Turn on scale. 00000 00000 00000 Press [CAL] , then release the key. Press and hold the T key, then press the F key, then release both keys.