Mb-Re Series - ISHIDA NOVA-II Setup Manual

Mtx series; mtb series; mte series; cws series; ncs series; qbx series; qb series; mb series; mb-re series; ib series; sb series;
Table of Contents

Advertisement

3.3

MB-RE Series

Follow the steps in the table below to set up and calibrate the MB-RE Series scale.
Step
1
Press the S key and release. [SA CAL] appears in the display.
2
The display blinks [CAL. 0] which indicates zero adjustment is automatically performed.
3
Display changes to [CAL. on] which shows the mode advanced to span adjustment.
4
Turn the calibration knob counterclockwise to CAL until it stops (see figure 4.1 below). Now the calibration
weight is loaded. Display starts flashing to indicate span adjustment is being automatically performed.
5
Display changes to [CAL oFF].
6
Return calibration knob at WEIGH point by turning it gently until it stops with a slight sound. Calibration weight
inside has been unloaded to display [End].
7
When span calibration is complete, [0] is shown on the display.
NOTES:
Pressing any other key other than
Calibration knob should be turned gently, but don't stop until fully turned. At
it stops with a sound.
[WEIGH],
Keep calibration knob at
while it is at
[CAL]
The calibration will not advance after
windshield or a stable base.
If
[3-Err]
appears on the display after zero adjustment is performed, an object other than the calibration
weight is on the pan.
If
appears on the display after returning the knob to
[4-Err]
varies too much from that of the factory-set value.
18
Ishida Setup & Calibration Manual
ON/OFF
[WEIGH]
except during calibration.
.
[CAL.0]
Unit No. (mark)
MB-300RE
1 (g)
2 (ct)
3 (Lb/oz)
Figure 3-1. Calibration Knob
Action
will interrupt the operation and show
[CAL.oFF]
if affected by wind or external vibration. Provide a
[WEIGH]
MB-600RE
300
600
1500
3000
10
21
[StOP].
[CAL]
it stops lightly, at
appears if the power is turned on
, this indicates that the error at span