Ib Series - ISHIDA NOVA-II Setup Manual

Mtx series; mtb series; mte series; cws series; ncs series; qbx series; qb series; mb series; mb-re series; ib series; sb series;
Table of Contents

Advertisement

3.4

IB Series

Follow the steps below to set up and calibrate the Ishida IB Series scale.
CALIBRATION
Step
1
Turn on scale.
2
Press and hold the F key until the [CAL] message is displayed.
3
Press and hold the T key, press the F key, then release both keys.
4
After zeroing procedure has finished, the following message will be displayed. Set
standard weight in weigh pan. The scale automatically calibrates itself.
5
Scale automatically returns to original measurement mode.
NOTES:
The Ishida IB Series precision balance automatically calibrates itself.
Press the
key to cancel calibration procedures and return to the weigh display.
P
If
appears on the display, the weight exceeds the scale capacity.
[o-Err]
Standard calibration weight is
If
[1-Err]
appears on the display, the calibration weight is less than 50% of the standard.
If
[2-Err]
appears on the display, an object other than the calibration weight is on the pan or the display
tolerance (1%) is exceeded.
Action
capacity. When possible, calibrate using full capacity.
1/2
Display
0000 0000 0000
CAL
on 0
on F.S.
50000
Balances
19