Mb Series - ISHIDA NOVA-II Setup Manual

Mtx series; mtb series; mte series; cws series; ncs series; qbx series; qb series; mb series; mb-re series; ib series; sb series;
Table of Contents

Advertisement

3.2

MB Series

NOTE: MB Series scales cannot be calibrated. They are stamped or sealed by local weights/measurement offices. When
calibration is necessary for the MB, contact your local dealer.
Follow the steps in the table below to set up and calibrate the MB-E Series scale.
Step
1
Press and hold the F key until [Func] –> [CAL] appear in sequence, then release the key.
2
Press the ZERO key. While holding it down, press the F key, then release both keys.
The display blinks [on 0] to indicate zero adjustment is being performed. Verify that no load is on the weigh
platter.
3
After the MB-E performs zero adjustment, [on FS] automatically appears. Gently place the designated
calibration weight on the center weigh pan (see table below for designated calibration weights).
4
When calibration is complete, the scale automatically returns to the weighing mode.
NOTES:
In Step 2, if the
F
Standard calibration weight is
weight. See the table below for the unit system capacities.
If
appears on the display, the weight exceeds the scale capacity.
[o-Err]
If
[1-Err]
appears on the display, the calibration weight is less than 50% of the standard.
If
[2-Err]
appears on the display, the display tolerance (0.4%) is exceeded, or the scale may be defective.
Contact your Ishida representative.
Unit No. (mark)
MB-150E
1 (g)
150
2 (ct)
750
3 (Lb/Oz)
5.2
key is pressed prior to the
capacity. However, when possible, calibrate using a full capacity
1/2
MB-300E
MB-600E
300
1500
3000
10
Action
key, the scale will return to weighing mode.
ZERO
MB-1500E
600
1500
7500
21
52
MB-3000E
MB-6000E
3000
6000
15000
30000
6.6
13
Balances
17