Adjustment Routine For Bvc Professional With 2L Glass Bottle - vacuubrand BVC control Instruction Manual

Biochem-vacuucenter
Hide thumbs Also See for BVC control:
Table of Contents

Advertisement

Adjustment routine for BVC professional with 2l glass bottle

from software version V1.9, see rating plate
Adjustment routine for older software versions: See "Instructions for adjust-
ment of the liquid level sensor" on www.vacuubrand.com.
Attention: Do not interrupt the adjustment routine once you have
begun, even if there is an alarm. Complete the entire routine, and
repeat if necessary.
Adjustment routine may take up to 3 minutes.
During adjustment all keys are without function.
Switching off the BVC during adjustment routine terminate the ad-
justment. The level sensor retains the original adjustment values.
Action
Step 1
Touch and hold "2l Glass"
key.
Step 2
When the LED of the "2l
Glass" key flashes blue,
touch and hold the "bottle
change" key at the same
time.
Step 3
Release both "2l Glass"
key and "bottle change"
key.
page 44 of 76
Reaction
Case 1:
The LED of the "2l Glass"
bottle symbol flashes
blue.
Case 2 / false alarm:
The alert tone of the false
alarm is shut off.
The LED of the "2l Glass"
bottle symbol flashes
blue.
A one-time buzzer
sounds.
The pump stops.
After the buzzer, the LED
of the "bottle change" key
lights red.
The LED of the "2l Glass"
bottle symbol flashes
blue.
All LED´s of the suction
power display flash blue.
The LED of the "bottle
change" key lights red.
Meaning
Adjustment is possible by
continuing to hold the "2l
Glass" key, and proceed-
ing to step 2.
Adjustment starts.
Adjustment is running.