During Operation - vacuubrand BVC control Instruction Manual

Biochem-vacuucenter
Hide thumbs Also See for BVC control:
Table of Contents

Advertisement

During operation

1
➨ Attention: When removing the aspiration
controller VHC
Adopt suitable measures to prevent the re-
lease of dangerous, toxic, explosive, corro-
sive, noxious or polluting fluids, vapors and
gases. In such cases, use an appropriate
collecting and disposal system and take pro-
tective action for personnel, equipment and
environment.
max.
+ Max. ambient temperature: 40 °C
40°C
• Use the system only with the integrated hy-
drophobic filter to protect the vacuum sup-
ply from aspirated liquids and aerosols, and
to protect the environment/user from con-
tamination risk.
NOTICE
The BVC professional (with 4l polypropylene
2
bottle) is equipped with quick couplings at the
bottle head (1) and at the inlet of the pump (2).
Quick couplings for other versions are avail-
able as accessories. (See section "Accessories
- Spare parts".)
Quick coupling set (1): VHC
When disconnected, the collection bottle clos-
es vacuum tight.
Quick coupling set (2): Bottle to pump unit:
When disconnected, the collection bottle closes
vacuum tight.
page 29 of 76
liquid in the tube may leak!
pro
to bottle:
pro

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bvc control gBvc professionalBvc professional g

Table of Contents