Notes Regarding The Setting Of The Suction Power - vacuubrand BVC control Instruction Manual

Biochem-vacuucenter
Hide thumbs Also See for BVC control:
Table of Contents

Advertisement

Notes regarding the setting of the suction power

Depending on ambient atmospheric pressure (depending on altitude or
weather conditions) and on the vapor pressure of the media inside the
collection bottle, the setting of a high underpressure (e. g., 850 mbar, 8
LEDs flash) may lead to continuous operation of the pump. In this case,
a reduction of the suction power is recommended.
Example:
Ambient pressure in high altitudes or in a depression: 800 mbar.
Suction power set to 8 LED´s, 850 mbar underpressure relative to ambi-
ent pressure.
Result: Pump is continuously running, the set pressure difference is nev-
er achieved.
Recommendation: Reduce suction power.
Setting suction power
1 LED
2 LED´s
3 LED´s
4 LED´s
5 LED´s
6 LED´s
7 LED´s
8 LED´s
Use the keys to select the bottle size (2 liter glass bottle or
4 liter polypropylene bottle), and with it the activation of the
corresponding level sensor. To operate the key "2l Glass"
or "4l PP" touch the key > 1 second. The LED of the se-
lected bottle flashes blue.
Use the key to turn off the level sensor alarm and to start or stop
the pump during bottle change.
To operate the key "bottle change", touch the key below the LED
> 0.5 seconds.
page 26 of 76
Underpressure relative to atmospheric
pressure / ambient pressure
150 mbar
250 mbar
350 mbar
450 mbar
550 mbar
650 mbar
750 mbar
850 mbar