Safety during operation

➨ Avoid interactions of media in the collection bottle.
For example sodium hypochlorite (chlorine bleach)
Incompatible chemicals and
agents
Acids or acidic compounds
(e. g. hydrochloric acid,
aluminium chloride)
Ammonia containing compounds
(e. g. ammonium hydroxide,
quarternary ammonium salts)
Organic chemicals
(e. g. solvents, polymers,
amines, oils)
Metals (e. g. copper, iron)
Hydrogen peroxide
Reducing agents
(e. g. sodium thiosulfate)
Guanidine salts
(e. g. guanidine hydrochloride,
guanidine thiocyanate)
➨ Avoid the formation of dangerous reactions / gases in
page 11 of 76
Comply with material safety data sheets and notes on
safe use of the manufacturer.
Do not mix incompatible disinfectants and/or in-
compatible reagents / solvents or any unknown
substances.
the BVC, especially in the collection bottle. If this is im-
possible dispose of dangerous gases at the outlet of
the pump appropriately.
Possible results of mixing with
sodium hypochlorite (chlorine
bleach)
Release of chlorine gas
Formation of explosive compounds,
release of chlorine gas and other
hazardous gases
Formation of chlorinated organics,
release of chlorine gas and other
hazardous gases
Release of oxygen, overpressure,
rupture of a closed system
Production of heat, boiling
Release of toxic gases, e. g. chlo-
rine, chloramine, hydrogen cyanide