Summary of Contents for vacuubrand BioChem-VacuuCenter BVC 01
Page 1
page 1 of 31 Technology for Vacuum Systems Instructions for use BVC 01 BVC 21 NT BVC 21 NT VARIO BioChem-VacuuCenter Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully the validity of this document with respect to his product.
Page 2
2 of 31 Dear customer, Your VACUUBRAND diaphragm pumps should support you for a long time without trouble and with maximal power. Thanks to our long practical experience we have much information how you could en- sure powerful application and personal safety. Please read these instructions for use before the initial operation of your pump.
page 3 of 31 Contents Safety information! .....................4 General information ..........................4 Intended use ............................4 Setting up and installing the system ....................4 Ambient conditions ..........................5 Operating conditions ..........................5 Safety during operation ........................6 Maintenance and repair ........................8 Technical data ......................9 Wetted materials ..........................10 System parts ...........................
☞ Make sure that the individual components are only connected, combined and op- erated according to their design and as indicated in the instructions for use. Use only original VACUUBRAND accessories. ☞ Comply with notes on correct vacuum and electrical connections, see section ”Use and operation”.
page 5 of 31 ☞ Do not permit any uncontrolled pressurizing (e.g. make sure that an exhaust pipeline cannot become blocked). Risk of bursting! ☞ Due to the high compression ratio of the pumps, pressure at the outlet port might be generated being higher than the maximum permissible pressure compatible with the mechanical stability of the system.
page 6 of 31 ➨ The devices are not suitable to pump unstable substances and substances which react explosively under impact (mechanical stress) and/or when being exposed to elevated temperatures without air, self inflammable substances, substances which are inflammable without air and explosive substance •...
Page 7
page 7 of 31 • Use the system only in combination with the integrated hydrophobic filter (steriliz- able) to protect pump and environment (people). • Comply with applicable regulations when disposing of chemicals. Take into con- sideration that chemicals may be polluted. Take adequate precautions to protect people from the effects of dangerous substances (chemicals, thermal decompo- sition products of fluoroelastomers), wear appropriate safety-clothing and safety glasses.
page 8 of 31 Maintenance and repair Wear parts have to be replaced regularly. In case of normal wear the lifetime of NOTICE the diaphragms and valves is > 10000 operating hours. Bearings have a typical du- rability of 40000 h. Motor capacitors have a typical durability in the range of 10000 to 40000 h depending strongly on the operation conditions like ambient temperature, humidity or load.
page 9 of 31 Technical data Type BVC 21 NT VARIO BVC 21 NT BVC 01 Pump Pump Valve Maximum pumping speed 1.9/2.2* (ISO 21360) Ultimate vacuum (absolute) mbar Maximum permissible inlet pres- sure (absolute) Maximum permissible outlet pressure (absolute) Rated power 0.09 0.12...
page 11 of 31 System parts BioChem-VacuuCenter BVC 21 NT / BVC 21 NT VARIO (fig.: BVC 21 NT VARIO) filter connecting tube inlet diaphragm pump with coupling (double sided sealed if removed) blind cap (for further connections) screw cap coupling collecting bottle adjusting knob for...
Page 12
12 of 31 BioChem-VacuuCenter BVC 01 connecting tube filter blind cap coupling (for further connections) (connection VHC) screw cap collecting bottle pressure control device inlet isolation valve with coupling (double-sidedly sealed if removed) solenoid operated hose nozzle isolation valve (connection to vacuum supply) ➨...
page 13 of 31 Use and operation Installation • It is imperative to remove the transportation lock prior to use! Tilt device care- fully onto the side. Remove the four locking screws (wing screws) next to the fan guard at the base plate. Put device back to normal operating position and keep the locking screws.
Page 14
page 14 of 31 Attention: In case of supply voltage below 100V, the lock of the cutout might be restricted and the pump might restart on its own after sufficient cooling down. Take suitable precautions, if an automatic restart of the pump may lead to a critical dan- gerous situation.
page 15 of 31 Shutdown Short-term: NOTICE Has the pump been exposed to condensate? Allow the pump to continue to run at atmospheric pressure for a few minutes. Has the pump been exposed to media which may damage the pump materials or form deposits? Check and clean pump head if necessary.
page 16 of 31 VacuuHandControl VHC The instrument allows the aspiration of excess liquids. The VHC is supplied completely with three differ- ent adapters (included in delivery) to hold with pasteur pipettes, capillaries and pipette tips (100 µl). Before aspiration: Push back the locking ring from the control button.
Page 17
page 17 of 31 The hydrophobic filter retains water and aqueous solutions also in the form of aerosols. When using solvents or due to evaporation of water, condensate may be recovered in the separator. ☞ Remove separator and drain condensate in case. Important: Comply with applicable regulations when dispos- ing of solvents.
page 19 of 31 Troubleshooting Fault Possible cause Remedy ❑ Pump fails to start, sys- ➨ ✔ Mains not plugged in, failure Plug in mains plug. check tem becomes vented. of electrical supply? mains fuse. ➨ ✔ Motor overloaded? Allow motor to cool, identify cause of failure.
page 20 of 31 Replacing diaphragms and valves All bearings are encapsulated and are filled with long-life lubricant. Under normal ope- NOTICE rating conditions, the pump is maintenance free. The valves and diaphragms as well as the motor capacitors are wear parts. If the rated ultimate vacuum is no longer achieved or in case of increased noise level, the pump interior, the diaphragms and the valves must be cleaned and the diaphragms and valves must be checked for cracks or other damage.
page 21 of 31 Removing cover hood from base plate ➨ Switch device off and unplug mains cable. remove ➨ Uncouple hose between collecting bottle and pump at the adjusting quick coupling. Remove bottle from its holder. knob BVC 21 NT VARIO: ➨...
Page 22
page 22 of 31 ➨ To check valves use hex key to remove four socket head screws from pump head and remove upper housing (housing cover with housing cover insert), head cover and valves. ☞ Position the pumping unit onto its left side with the pump head at the top to facilitate maintenance.
➨ Optimum torque for the diaphragm support disc: 6 Nm. ☞ The optimum torque is achieved if the pointer in the handle of the VACUUBRAND face wrench shows to the longer marking line. Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully...
page 24 of 31 Assembling the pump head ➨ By turning eccentric bushing (front of connecting rod), bring connecting rod into a position in which diaphragm is in contact with housing and centred with respect to bore. ☞ Tilt pump carefully onto the left side so that the diaphragm is on top.
page 25 of 31 Mounting the cover hood ➨ Position cover hood on the pump. ☞ Pay attention that the metallic rear panel runs without can- ting in the slots of the cover hood. Slide on cover hood completely. ☞ Do not cant the cover at the ducts of the quick coupling or the adjusting knob.
page 26 of 31 Assembling of components Replacing the aspiration tube at the VHC Ensure that there is no liquid in the tube, risk of contamina- tion! Remove adapter from the fixing clip. While pressing the con- trol button, pull out the aspiration tube. Assembling new aspiration tube: While pressing the control button, insert the new aspiration tube between thumb wheel and roller.
page 27 of 31 Cleaning and decontaminating Autoclaving: The aspiration tube and the pipettes of the VacuuHandControl VHC, the collecting bottle with bottle head and screw cap and the filter are designated for steam sterilisation at 121°C and 2 bar absolute (1 bar overpressure).
page 28 of 31 Notes on return to the factory Repair - return - DKD calibration NOTICE Safety and health of our staff, laws and regulations regarding the handling of dan- gerous goods, occupational health and safety regulations and regulations regarding safe disposal of waste require that for all pumps and other products the “Health and safety clearance form“...